From cardiac cells to the ECG
The heart generates the electrical source. The recorder measures voltage differences at the skin.
Ventricular activation
Leftward, inferior components · prescribed source
The His–Purkinje system recruits ventricular muscle at multiple sites. The visible boundary is a teaching schematic; the arrow summarizes a prescribed net source, not the path of one ion. Source ↗
Cutaway schematic; chamber geometry and activation boundary are illustrative. The arrow’s chosen origin is a drawing convention.Potential at the electrodes
Lead II
The recorder subtracts the two electrode potentials. It does not create the source dipole or pull ions through the heart.
Example acquisition electronics ↗This ideal measurement draws no input current. Real front ends have finite input impedance and may use bias or common-mode feedback circuits; those are not the mechanism generating the ECG.Synthetic lead II
The cursor, source arrow, body field and electrode readings use the same time.
Scrollable cycle. Use the time slider to follow the cursor.
Grid spacing: 50 ms horizontally · 0.5 mV vertically. Prescribed P/QRS/T source functions, not a cellular simulation or measured heartbeat. Changing a lead or reversing inputs changes the measurement; it does not change cardiac activity.
The violet arrow is drawn near the heart for reference. Its length and direction represent an equivalent current dipole, p, summarizing the distributed source in this model. It is not emitted by an electrode. The dashed lead axis represents measurement sensitivity. Neither arrow is a stream of ions.
Why a moving boundary makes a signed voltage
Change the process and its direction. The lead’s positive input stays on the right.
− input at left · + input at right. Scroll horizontally to inspect the full circuit.
Upward deflection
Activation creates a spatial gradient between activated and resting tissue. In this simple geometry its equivalent dipole points in the direction of activation.
Opposite process + opposite direction = same sign. This explains how a T wave can be upright after an upright QRS. It does not establish a particular person’s recovery sequence.
Boundary-polarity derivation · Bioelectromagnetism §15.2 ↗What crosses the membrane? Why are the cells not simply charged beads?
Scroll horizontally to compare Na⁺, Ca²⁺ and K⁺ pathways.
Working ventricular muscle, simplified. Voltage-gated channels conduct ions down electrochemical gradients. Sodium influx drives the fast upstroke; calcium and potassium currents contribute to the plateau; net outward current produces repolarization. Nodal cells have different currents and timing. Adjacent cells are electrically coupled through gap junctions; activation travels as local current recruits nearby membrane, not as one ion crossing the entire heart. Human ventricular ionic model ↗
Membrane voltage equals intracellular minus extracellular potential.Vm, φi and φe are potentials in mV. The approximate −90 and +20 mV labels illustrate resting and activated membrane states; they are neither simulated patient values nor the ECG amplitude. Most bulk fluid remains nearly electrically neutral. Small charge separations across membranes and associated current loops matter; a surface electrode samples extracellular potential after volume conduction.
Inspect the equations, units, assumptions and source evidence
One forward calculation drives the diagrams
Potential equals the dipole moment dotted with the source-to-electrode position, divided by four pi times conductivity times distance cubed. Lead voltage equals positive-input potential minus negative-input potential.Exact for the selected idealization. A quasistatic point current dipole in an infinite homogeneous isotropic conductor, with φ → 0 at infinity. This is a closed-form calculation, not a numerical torso solution. Each field tile and electrode uses this equation. The trace repeats the electrode subtraction across time. The chamber boundary and cellular illustrations explain mechanisms; they are not inputs to this calculation. Source-field derivation ↗
- p(t) · A m
- Current dipole moment; prescribed synthetic source. Current times source–sink separation, directed from sink to source. It is not a charge dipole in C m.
- r · m
- Position from source origin to field point. Frontal slice z = 0; axes point to the patient’s left and inferiorly.
- σ · S/m
- 0.2 S/m, a chosen illustrative uniform conductivity; not fitted to an individual torso.
- φ · V
- Potential relative to infinity. Labels convert volts to mV. Only differences enter the lead reading.
- R · m
- 0.25 m, chosen equal electrode radius. RA = (−√3 R/2, −R/2), LA = (√3 R/2, −R/2), LL = (0, R).
- p scale
- The existing synthetic source components are multiplied by 4πσR²/(1000√3) A m per numerical source unit, explicitly chosen to retain the teaching waveform’s mV scale. No physiological fit is implied.
L · V/(A m) is measurement sensitivity at the fixed source. In this particular centered equilateral geometry, it is parallel to the dashed negative-to-positive lead axis. A general torso lead field varies spatially and need not follow a straight wire between electrodes. A perpendicular dipole can give zero in one lead while remaining visible in another. The potential-difference definition is exact; a single cardiac vector describing a distributed heart is an approximation. Distributed sources and volume-conductor theory · Geselowitz (1967) ↗
What this explanation does and does not establish
- Cellular mechanism: the ionic close-up summarizes working ventricular muscle. It omits detailed channel kinetics, calcium handling, fiber anisotropy and nodal differences. ten Tusscher et al. (2004), human ventricular model ↗
- Activation: Durrer et al. mapped activation in seven isolated human hearts. That supports multiple ventricular breakthrough sites and organized activation; it does not validate this illustrative chamber shape or source waveform. Durrer et al. (1970) ↗
- Recovery: the tissue-boundary sign exercise is a one-dimensional explanation. Recovery timing follows regional action-potential duration and activation time; it is not generally a traveling excitation wave. A real T wave cannot be explained by reversing one arrow alone. Mapping in nine perfused pig hearts found contributions from multiple anatomical recovery gradients, not only one transmural direction. Meijborg et al. (2014), animal preparation ↗ · Local polarity derivation ↗
- Inference limit: a single voltage difference cannot uniquely recover a spatial source vector. Real torso conductivity, lungs, blood, geometry, electrode contact, filtering and distributed sources affect the recording. This view is a forward teaching model and never fits an uploaded record.
- Instrumentation: surface ECG acquisition senses differential potentials with high input impedance. Some electronics use right-leg common-mode feedback, input bias return or lead-off detection currents. Those circuits are distinct from pacing and do not explain the origin of normal cardiac excitation. ADS129x acquisition datasheet ↗